The Compton scattering of the 662 keV gamma rays from the decay of Cs137 is measured using a Sodium Iodide detector. The scattered energy and the differential cross section are both measured as a function of scattering angle, and the results are compared to the full relativistic quantum theory of radiation.
Theory
The Compton effect is the elastic scattering of photons from electrons. As a reaction, the process is:
|
Since this is a two body elastic scattering process, the angle of the scattered photon is completely correlated with the energy of the scattered photon by energy and momentum conservation. This relation is usually written as:
| (1) |
The energy of a photon is related to its frequency and wavelength as:
| (2) |
| (3) |
| (4) |
The quantity h/mc in Eq. () is called the Compton wavelength and has the value:
|
As an example, in this experiment gamma rays from Cs137 are scattered from an aluminum target; since E = 0.662 MeV, we have g = 1.29 so that gamma rays scattered at q = 180° will have E¢ = E/3.6, which is less than 1/3 of their original energy. It thus becomes quite easy to observe the Compton energy shift. This would not be the case for X-ray energies.
Another useful kinematic relation is the electron scattering angle in terms of the photon scattering angle:
|
The above kinematic relations as well as the following discussion on cross sections may be found in Melissimos pp. 252-265 which is required reading for this experiment.
When hn << mc2 the probability for Compton scattering can be regarded as a classical process and is given by the Thompson cross section which is the classical limit of the exact Compton scattering cross section formula.
| (5) |
| (6) |
the electron, although free, is assumed not to recoil;
the treatment is nonrelativistic;
quantum effects are not taken into account.
The problem was solved by Klein and Nishina in 1928 giving the correct quantum-mechanical calculation for Compton scattering, the so called Klein-Nishima formula:
|
|
The following table and graph show the Klein-Nishina cross section as a function of photon scattering angle. The table is for the .662 MeV gamma ray energy of Cs137, while Fig. shows the angular distribution for a range of incident photon energies.
q | cosq | Classical E | Relativistic E | Klein-Nishima |
(Degrees) | (keV) | (keV) | (10-30 m2) | |
0 | 1.00000 | 662.00 | 662.00 | 7.941 |
5 | .99619 | 658.74 | 658.75 | 7.834 |
10 | .98481 | 649.06 | 649.22 | 7.524 |
15 | .96593 | 633.24 | 634.01 | 7.047 |
20 | .93969 | 611.73 | 614.03 | 6.452 |
25 | .90631 | 585.20 | 590.35 | 5.793 |
30 | .86603 | 554.45 | 564.09 | 5.119 |
35 | .81915 | 520.46 | 536.34 | 4.471 |
40 | .76604 | 484.32 | 508.02 | 3.875 |
45 | .70711 | 447.17 | 479.90 | 3.348 |
50 | .64279 | 410.13 | 452.57 | 2.894 |
55 | .57358 | 374.23 | 426.43 | 2.513 |
60 | .50000 | 340.31 | 401.76 | 2.200 |
65 | .42262 | 308.95 | 378.72 | 1.948 |
70 | .34202 | 280.49 | 357.37 | 1.746 |
75 | .25882 | 255.06 | 337.72 | 1.589 |
80 | .17365 | 232.58 | 319.72 | 1.467 |
85 | .08716 | 212.88 | 303.31 | 1.374 |
90 | .00000 | 195.71 | 288.39 | 1.304 |
95 | -.08716 | 180.79 | 274.87 | 1.254 |
100 | -.17365 | 167.86 | 262.65 | 1.217 |
105 | -.25882 | 156.65 | 251.63 | 1.192 |
110 | -.34202 | 146.95 | 241.73 | 1.176 |
115 | -.42262 | 138.55 | 232.85 | 1.166 |
120 | -.50000 | 131.29 | 224.92 | 1.161 |
125 | -.57358 | 125.01 | 217.87 | 1.160 |
130 | -.64279 | 119.60 | 211.62 | 1.161 |
135 | -.70711 | 114.95 | 206.13 | 1.164 |
140 | -.76604 | 110.99 | 201.34 | 1.168 |
145 | -.81915 | 107.63 | 197.22 | 1.172 |
150 | -.86603 | 104.83 | 193.71 | 1.176 |
155 | -.90631 | 102.53 | 190.80 | 1.181 |
160 | -.93969 | 100.69 | 188.45 | 1.184 |
165 | -.96593 | 99.29 | 186.64 | 1.187 |
170 | -.98481 | 98.31 | 185.37 | 1.190 |
175 | -.99619 | 97.72 | 184.60 | 1.191 |
180 | -1.00000 | 97.53 | 184.35 | 1.191 |
Radiation Units
Milliroentgen per hour (mR/h) are units of radiation exposure. Exposure indicates the production of ions in a material by radiation, and it is defined as the amount of ionization produced in a unit mass of dry air at standard temperature and pressure. The roentgen is the conventional unit for exposure, where
|
Radiation safety standards are expressed in units of roentgen equivalent mammal per year (rem/yr). We now relate roentgens to rems via the rad unit and the RBE or QF, discussed below.
The absorbed dose is the radiation energy absorbed per kilogram of absorbing material. It is measured in rads, where
absorbed dose unit: 1 rad = 0.01 joule per kilogram.
In animal tissue it takes about 30 eV or 4.8×10-18 J to produce an ion pair, and assuming the magnitude of the charge of each ion is 1.60×10-19C, then for animal tissue
|
The effects of radiation on biological systems depend on the type of radiation and its energy. The relative biological effectiveness (RBE) or quality factor (QF) of a particular radiation is defined by comparing its effects to those of a standard kind of radiation, which is usually taken to be 200-keV x rays. The RBE or QF is the ratio of the dose in rads of a particular kind of radiation to a 1-rad dose of 200-keV x rays, where the particular radiation produces the same biological effect as the x rays. Note that RBE or QF is dimensionless.
|
The rem is defined as
|
Radiation standards adopted by the United States Government are the following:
For the general population living near a nuclear facility: 0.5 rem/yr.
For worldwide population: 5 rem total up to age 30 (0.17 rem/yr), in addition to natural background radiation, which is about the same intensity. Primary concern is for genetic damage. It is estimated, rather uncertainly, that 0.17 rem/yr may produce 5000 extra deaths and 5000 birth defects in the United States per year.
Shielding Calculations
The calculations consist of two parts. First we calculate the attenuation of gammas which is desirable. Then we calculate the thickness of lead to give this attenuation.
|
|
Hence we have
|
79.7 rem/year (if no shielding)
|
Gamma Source
The gamma source is Cs137. The present source had a nominal strength of 100 millicuries when it was purchased in 1975, although the actual strength appears to have been somewhat smaller. A careful measurement of the source strength in May, 1991, yielded a value of 42±2 millicuries.
This is a strong source and so there are 3 very important rules. The decay scheme of Cs137 is shown in Fig. 3. The details of the source are shown in Fig. 4.
The container barrel points towards a concrete wall. Do not try to move the container or table.
Do not place any part of your body within 15° of the barrel axis unless the stopper is in the position to block the barrel.
The lead container is designed conservatively so that the dose you would receive if you leaned against the container for 16 hours each month would be still less than 10% of the allowable dose for students and 2% of the allowable dose for industrial workers. The dose is approximately 1% of the natural background dose. As a result, the lead container weighs about 540 lbs. or 250 kg.
In practice, of course, you will spend less time than 16 hours and will be at a distance which reduces the background intensity 50 fold.
Source Container and Barrel
The container is made of lead and so the 662 KeV gammas have a nearly equal probability of interacting via the Photo-Electric effect and via the Compton Effect. The most important angle of scattering is in the forward ±45 ° cone because
As we require a jet of well collimated 662 KeV gammas with as little contamination as possible of gammas with lower energies, the container barrel has been designed to greatly reduce the chance of gammas escaping after a forward scatter (q < 90°).
If a 662 KeV photon is scattered through more than 45° then it is likely to be absorbed by a small thickness of lead.
|
Angle of Scattering | Energy | l in Lead |
0° | 662 keV | 0.88 cm |
45° | 479 keV | 0.57 cm |
90° | 288 keV | 0.22 cm |
135° | 206 keV | 0.11 cm |
180° | 184 keV | 0.09 cm |
The gammas which do not pass out through the aperture are incident upon the surfaces A, B, C, D. (See Fig. 5.)
A and B A few gammas will Compton Scatter to give 184 MeV gammas which scatter to the right. However, the 662 gammas will have penetrated about 0.88 cm into the lead and so only a few of the 184 MeV gammas will escape from inside the AB surfaces. Some gammas will interact via the photoelectric effect. Both will cause 75 keV X-rays (Lead K shell) but again only a few X-rays will escape from inside the AB surfaces.
C The aperture is designed so that scattered gammas and X-rays cannot escape directly from C.
D Again scattered gammas and X-rays cannot escape directly. Although the Compton scatters of 90° to 45° may have enough energy for a second Compton scatter on the opposite surface E, the distance they must travel from the interaction point in D to leave D is greater than the penetration into D. Note that a few gammas will Compton Scatter near the DE corner and will escape into the main beam. Obviously if we had a ring of dense HIGH Z material such as uranium, at the corner, this background could be reduced. See Fig. 6.
E This is a conical surface designed so that it cannot be directly illuminated by gammas from the source.
F The forward outer surface is positioned so that 662 keV gammas leaving the source must travel through a distance of 12l, or 10.56 cm. The attenuation factor is then approximately e-12 = 6 ×10-6. The total gammas intensity is greater since the scattered gammas are not all absorbed. The ``Build Up Factor" for gammas of 1/2 MeV in lead is about 2 and so the overall attentuation is about 12×10-6.
G The backward outer surface has an extra 2l added as an extra (and unnecessary) safety factor.
|
|
The approximate angles are:
Angle of scatterer subtended at the source: ±5°
Angle of Uniform g jet from the source: ±10°
Angle of outside of g jet from the source: ±20°
Angle of scintillator subtended at the scatterer: ±8°
Minimum angle of scattering which is outside the jet: 15°.
Tapered Plug
This is intended to block the beam. The taper is made from brass and so must be longer than if it had been made from lead. We except a few scattered gammas to sneak along the taper since the fit cannot be perfect. For this reason an extra 3.8 cm of steel is used to block those escaping along the taper.
Detector (Harshaw Chemical Type 858 Serial 6V230
with voltage divider)
This is a NaI crystal scintillation cylinder with a diameter of 2 inches and a thickness of 2 inches. The crystal is hermetically sealed and in good optical contact with the photocathode of an RCA 6342A phototube with 10 stages. The Harshaw Quality Assurance Report states that the detector has a resolution when measuring 662 keV gammas of
|
The photomultiplier requires a positive high voltage via the white cable and High Voltage BNC connector. The dynode potentials are controlled by the built in voltage divider with a total resistance of 6.2 Mohms. The anode is at the positive high voltage and so a 1 nf capacitor is used to pick off the 1 msec 100 millivolt negative signals. The maximum HV rating is +1500V but the experiment needs only ~ 1100-1200 V.
The NaI-PM assembly is mounted inside a tapered lead cylinder. The lead weighs about 23 kg and is intended to serve as a collimator so that only gamma rays from the target region are detected. See Fig. 8.
The NaI crystal diameter is 2 inches but only 1 3/4 inches are exposed to the gammas to give the greatest probability of full absorption of each gamma entering the NaI.
The lead cylinder is mounted on rollers and rotates about a center which can support a scatterer.
Scatterer
The scatterer is an aluminum disc (1.27 cm thick, 3 cm diameter). The Compton effect depends only on the electrons but does assume that the electrons are loosely held. For this reason, the Z of the scatterer should not be too high. The K shell electrons in aluminum have a level of -1560 eV and the L shell electrons have levels of -118, -74 and -73 eV.
The scatterer was chosen as a disc so that the probability of multiple scattering may be minimized. The scatterer is mounted on rigid foam plastic so that the support will scatter few gammas.
High Voltage Supply
This is regulated and can supply up to 2.5 mA at +2.5 kV. This experiment requires about 190 mA at +1200 V. The gain G of the photomultiplier depends on the voltage, but the gain variation with voltage is approximately DG/G = n DV/V where n = 10 is the number of dynodes. A typical gain for our operating conditions is G @ 104. Consequently if we wish to measure the energy of scattered gammas to ±1%, then the HV must be regulated to better than ±(1/4)%.
Amplifier
An ORTEC Model 575 Amplifier is used to amplify the output of the sodium-iodide detector to match the 0-10 V full scale range of the multi-channel analyzer. The gain is normally set so that the Cs137 peak is in about channel 800.
Multi-Channel Analyzer (Ortec Model 916)
This is a 2048 channel ADC unit mounted inside the IBM PC with the input connector on the back of the PC. The program to run it is called by typing ``MCA" from anywhere inside the DOS operating system. Consult the instructor for details on running the associated software.
Procedure
Make sure the plug is back in the source.
At each angle take a run with and without the Al target. These runs do not have to be for equal times if they are correctly normalized.
The MCA software may be used to subtract the appropriately normalized background as you go along and to place regions of interest (ROI) about the detected gamma ray peaks. For each ROI, the software can calculate the center of the peak, the full width half maximum (FWHM) and the net area under the peak.
Spectra may be stored on the computer's hard disk (but please delete them sometime after you have completed all of your data analysis). You also have the option of printing out the spectrum as seen on the screen (PRT SCRN) or printing out the counts vs channel number by printing the file from within the MCA program.
For each spectrum, identify the channel number of the peak of the Compton scattering. Use the plot of (2) above to compute a scattered energy E2 for each angle. For each spectrum, obtain the yield of scattered photons. This may be done either of two ways: (a) accept all counts above noise level and apply a correction for the efficiency of the crystal, or (b) integrate the counts in the photopeak only and apply a correction for the ``photofraction" as well as for the crystal efficiency. For the latter method, it is probably best to sum the channels that fall within the FWHM of the detected peak and correct back to the full area. For a Gaussian shaped peak, the counts within the FWHM region correspond approximately to [1/ 1.314] of the total area. Efficiency and photofraction data for a 2¢¢diam.×2¢¢ thick NaI crystal are included as figures 10 and 11.
|
|
|
|
where I is the gammma intensity at the scatterer, N is the number of electrons in the scatterer, and G(q) is the rate of scattered gammas into the NaI at angle q.
Use the plot below (Fig. 10) for the efficiency. The differential cross section at each angle q is then
|